METHOD 6241.2
March 22, 2001
INFRARED REFLECTANCE FROM SPECTROPHOTOMETRIC DATA
1. SCOPE
1.1 Scope. This method is designed to provide a measure of infrared reflectance. The effective
infrared reflectance of a specimen for a specified wavelength interval and for specified geometric
conditions is dependent upon the spectral composition of the incident energy, the spectral
sensitivity of the receiver, and the spectral reflectance of the specimen. This method should be the
basic method for the measurement of infrared reflectance and in case of dispute; the results obtained
under this method should govern.
2. APPARATUS
2.1 Spectrophotometer. Use a spectrophotometer that determines the spectral characteristics of
the test specimen within the applicable limits of the infrared spectrum. Use a photoelectric type
apparatus suitable for determinations throughout the applicable wavelength interval. Design the
apparatus to minimize the effect of stray energy by introducing stray energy filters or use double
dispersion prisms (see 6.1). Adjust the direction of irradiation to be nearly perpendicular to the
irradiated face of the test specimen and view hemispherical or the converse. Use split widths of the
monochromator that are sufficiently narrow throughout the applicable infrared spectrum to insure
that the spectral characteristics of the specimen are not appreciably altered when narrower slit
widths are used.
2.2 Standard. Smoked or pressed MgO or pressed BaSO4 prepared according to Recommended
Practice for Preparation of Reference White Reflectance Standards, ASTM E259.
3. REAGENTS
3.1 Reagents. None required.
4. PROCEDURE
4.1 Irradiating and viewing. If glossy samples are to be measured, the specular components of
reflected energy shall be excluded wherever possible.
4.2 Test specimen. The area of the test specimen viewed shall be sufficiently large to be
representative of the test specimen, textured specimen in particular.
4.3 Calibrations. The calibration of the photometer scale shall be carefully checked at reasonable
time intervals in a manner to insure accuracy over the entire range. Glass filters shall be used for
this purpose. The wavelength calibration shall be checked frequently to insure accuracy. Spectrum
lamps shall be used for this purpose.