METHOD 6241.2
March 22, 2001
4.5 Calculation of reflectance from spectrophotometric data. Tabulate spectral reflectance data
for the specimen for the 30 wavelengths listed in table I and the 10 wavelengths in table II.
Calculate the average of the 30 spectral reflectance values in table I and the 10 values in table II.
4.6 Report. Report the average infrared reflectance of the specimen for the wavelength intervals
specified in tables I and II. If required, report the spectral reflectance of the test specimen either as a
plotted curve or as tabulated data. Report the method of specimen preparation, if any. Report the
reference standard and spectrophotometer used for the measurements.
5. PRECISION
5.1 Data. No data.
6. NOTES
6.1 Notes. The General Electric (Hardy) recording spectrophotometer, the Beckman Model DK-2A
spectrophotometer, and Cary Model 14 spectrophotometer should be used. The specular component
should be excluded. The Beckman DU spectrophotometer does not use hemispherical viewing and,
therefore, does not meet the geometric requirements of the method, but it should be used if no other
instrument is available.
FED. TEST METHOD STD. 141D
3